Table 4:

Cutoff values of 10 significant TDC parameters for detecting stenosis flow with optimized sensitivity and specificity

ROI VariableAUCP ValueCutoff ValueSensitivitySpecificity
rTmax
    M20.638 (0.555–0.722).0020.41583.6%37.6%
    PV0.615 (0.53–0.701).0114.90555.2%63.4%
    SSS0.655 (0.57–0.74)a.0016.89555.2%73.7%
    M10.663 (0.58–0.746)a.0010.4581.0%58.2%
    SS0.582 (0.495–0.669).0457.4273.5%91.2%
    JV0.589 (0.503–0.676).0457.7763.7%84.9%
MS
    M20.606 (0.517–0.695).02511.589.6%21.8%
    M10.689 (0.608–0.77)a.041378.366.7%60.0%
    SSS0.636 (0.547–0.725).00290.555.1%75.5%
FWHM
    M20.609 (0.52–0.699).0133.4557.1%67.7%
    M10.679 (0.59–0.77)a.0013.4057.1%78.4%
  • Note:—SS indicates sigmoid sinus; JV, internal jugular vein.

  • a The best 4 parameters for detecting stenotic flow.