Table 3:

Performance comparison of SPACE versus DANTE-CAIPI-SPACE+DL, values expressed as median (1st quartile, 3rd quartile). P values were reported to test the median of the differences between paired observations by bootstrap. Effect sizes were quantified by using Cliff's δ

MetricSPACEDANTE-CAIPI-SPACE+DLCliff’s δP Value
QualitativeImage quality3 (3, 3) 2 (2, 3) 0.37.35
SNR2 (1, 2)2 (1, 2)–0.00941
Artifacts4 (3, 4)3 (2, 4)0.4.079
Arterial blood suppression3 (3, 4)1 (1, 1.75)0.88<.001
Venous blood suppression3 (3, 4)1 (1, 2)0.87<.001
Lesion assessment2 (1, 2)2 (1, 2)–0.021
QuantitativeWall SNR10.82 (6.22, 16.43)9.71 (6.99, 15.53)0.017.90
Wall CNR5.21 (2.99, 9.49)4.24 (2.10, 7.65)0.13.33