PT - JOURNAL ARTICLE AU - Shellock, Frank G. AU - Gounis, Matthew AU - Wakhloo, Ajay TI - Detachable Coil for Cerebral Aneurysms: <em>In Vitro</em> Evaluation of Magnetic Field Interactions, Heating, and Artifacts at 3T DP - 2005 Feb 01 TA - American Journal of Neuroradiology PG - 363--366 VI - 26 IP - 2 4099 - http://www.ajnr.org/content/26/2/363.short 4100 - http://www.ajnr.org/content/26/2/363.full SO - Am. J. Neuroradiol.2005 Feb 01; 26 AB - Summary: A detachable coil was assessed for magnetic field interactions, heating (body RF coil, SAR of 2.0-W/kg), and artifacts at 3-Tesla. The coil showed no magnetic field interactions and heating was negligible (+0.3°C). Therefore, this coil is safe for a patient undergoing MR imaging at 3T or less. While artifacts may impair the ability to properly visualize anatomy in proximity to this implant, careful selection of imaging parameters can mitigate this problem.